Production » Dosimeters » X-RAY RADIATION DOSIMETER AT-1103M
X-RAY RADIATION DOSIMETER AT-1103M
Unique high-sensitive instrument to control dose burden of crystalline lens, mucous membranes and skin. It measures directional dose equivalent rate of continuous x-ray radiation with the energy from 5 keV
Features
- X-ray source search |
- Fast response to radiation background change |
- Audible and visual alarm at threshold exceeding |
- Large digital backlit LCD with analog scale |
- Keeping up to 100 measurement results in the dosimeter nonvolatile memory |
- Built-in LED stabilization system providing no need to use a reference source |
- 256-channel MCA |
- Logging and transferring measurement results to PC |
- Dust and splash-proof case |
- Three power types |
Application
- Control of acceptable low-energy and intensity x-ray levels from video monitors, night vision equipment, oscillographs, TV receivers, microwave generators, ion implantation devices, inspection and medical sets |
- Certification testing of instruments and equipment containing unused x-ray radiation sources, safeguards effectiveness control |
- Dosimetry contamination control of radioisotopes 55Fe, 239Pu, 129I, 241Am etc. |
The dosimeter has scintillation NaI (Tl), ø 9x2 mm, with beryllium window as a x-ray radiation detector. The measuring method of directional dose equivalent rate is based on measuring the instrument spectrum and its interval standardized weighing per a dose rate unit. The relevant energy response correction for the count rate mode is also provided.
Specification
Detector | NaI (Tl) - ø 9x2 mm with a beryllium window |
Directional dose equivalent rate measuring range | 0.05-100 µSv/h |
Directional dose equivalent measuring range | 0.05 µSv - 5 mSv |
Intrinsic measurement error | not more than ± 15 % |
Energy range | 5 - 160 keV |
Energy sensitivity response relatively 241Am | |
5 - 60 keV | ± 35 % |
60 - 160 keV | ± 30 % |
Calibration error | |
57Сo, 109Cd, 55Fe, 241Am | not more than ± 5 % |
Maximum input statistical load | 6·104 с-1 |
Operation mode setup time | not more than 5 min |
Detecting count rate range | 0.01-6· 104 с-1 |
Detectable activity of 241Am at 0.5 m for 1-2 s | 1000 kBq (27 ³Ci) |
Sensitivity on 241Am | 400 cps/µSv·h-1 |
Operating temperature range | 0 - +40°С |
Relative humidity +350С |
up to 90 % |
Protection class | IP54 |
Power requirements | |
built-in Ni-Mh accumulator unit | 6 V |
AC mains, 50 Hz | 220 V |
DC supply | 12 V |
Continuous operation time from fully charged accumulator unit |
not less than 24 h |
Radio disturbance | |
EN 55022:2006 | |
Electromagnetic compatibility | |
IEC 61326-1:2005 | |
Weight | 0.8 kg |
Dimensions | 233х85х67 mm |
Complete set: dosimeter, AC adapter, hand strap, handle, holster and Manual.
Telescopic bar 1.1 m, cable to connect PC and applied software, cable for DC supply, packing case are options and they are supplied on additional order.
The x-ray radiation dosimeter AT1103M has pattern approval certificates of Republic of Belarus, Russian Federation, Ukraine, Kazakhstan and Uzbekistan.
It complies with IEC 60846 International standard requirements.