Production » Spectrometers and radiometers of radiation » Gamma and beta spectrometer - radiometer MKGB-01 RADEK

Gamma and beta spectrometer - radiometer MKGB-01 RADEK

The spectrometric complex consists of:

  • Detection units BDEG-80, BDEG-63, BDEG-150, BDEG-K, BDEB-60 (on customer choice)
  • Analog-to-digital converter (analyser) MD-198
  • Controlled power supply of high voltage HV, power supply of low voltage LV
  • Power supplies and gain units for BDEG-K
  • Low-background chambers with shielding for units BDEG, BDEG-K and BDEB (on customer choice)
  • Set of calibrating point sources of ionizing radiation (on customer request)
  • Program for full spectrometric analysis "ASW"
  • Personal computer IBM-PC of any configuration (on customer choice)
  • Connecting cables

    The following units can be included in the consist of spectrometric complex on separate order:
  • Printing complex (Automation system of counting samples accounting and marking) Shtrih-3
  • Sample changer for beta and gamma tracts

Distinctive features

  • Simultaneous and selective measurement of beta and gamma radiation
  • Software in operating sistem Windows 98/ME/2000/XP/Vista/7
  • Visualization of spectrum set process
  • Conducting of computations during the spectrum set process
  • Measurements automation

Spectrometer-radiometer MKGB-01 RADEK is registered in the State Register of Measuring Instruments of Russian Federation under № 21730-01. Pattern Approval Certificate of Measuring Instruments RU.C.38.001.A N 10702.

 

 

 


Main characteristics (in pdf):

Detection units of gamma-radiation

- BDEG-80
scintillation on the base of monocrystal NaJ(Tl) for registration of gamma-radiation, with dimensions, mm

80х80

- BDEG-63
scintillation on the base of monocrystal NaJ(Tl) for registration of gamma-radiation, with dimensions, mm

63х63

- BDEG-150
scintillation on the base of monocrystal NaJ(Tl) for registration of gamma-radiation, with dimensions, mm

150х100

- BDEG-K
semiconductor germanium-lithium or HPGe for registration of gamma-radiation

- - with  sensitive surface area, mm2

no less than 100

- - with detection efficiency of gamma-radiation
on line 1332.5 keV of 60Со radionuclide with
point source position on
distance 20 cm from detector surface, relatively
to detector NaJ(Tl) with dimensions 76,2х76,2 mm, %

no less than 10

Detection units of beta-radiation

- BDEB-60
scintillating plastic for registration of beta-radiation, with dimensions, mm

60х12

Detectors power supplies

- Controlled power supply of high voltage HV,
power supply of low voltage LV

- BNV-31 or its analogue for BDEG-K

Analog-to-digital converter (analyser) MD-129, MD-198, MD-208

- with number of channels

1024, 2048,4096,8192

- connection type to IBM PC

LPT (MD-129), USB (MD-198), BlueTooth(MD-208)

Range of registered energies, keV

- of gamma-radiation

50 - 3000

- of beta-radiation

100 - 3000

Integral nonlinearity, %

- for scintillation gamma-spectrometer

no more than 1

- for scintillation beta-spectrometer

no more than 2

- for semiconductor gamma-spectrometer

no more than 0,05

Energy resolution

- with detection unit BDEG-80 (BDEG-63) on gamma-line
661,7 keV of 137Cs radionuclide, %

no more than 9

- with detection unit BDEG-150 on gamma-line
661,7 keV of 137Cs radionuclide, %

no more than 13

- with detection unit BDEG-K on gamma-line 1332,5 keV of 60Со radionuclide, keV

no more than 3,0

- with detection unit BDEB on conversion electrons peak with energy 624 keV of 137Cs radionuclide, %

no more than 15

Maximum input statistical load, imp-1

no less than 50000

Continuous operation time, h

24

Instability,
(change of sensitivity) per 8 hours of working, %

no more than ±3

Operation mode setup time, min

no more than 30

Measured radionuclides

- with unit BDEG-K

137Cs, 106Ru-106Rh, 134Cs, 144Ce-144Pr, 125Sb, 241Am, 226Ra, 232Th, 40K and etc.

- with unit BDEG-80 (BDEG-63, BDEG-150)

137Cs, 134Cs, 238U (226Ra), 232Th, 40K, 131I, 60Co

- with unit BDEB

90Sr-90Y, 137Cs

Minimum detectable activity with measuring time 1 hour, Bq

- 137Cs

no more than 2

- 90Sr+90Y

1

- 226Ra

no more than 8

- 232Th

no more than 5

- 40K

no more than 40

- other gamma-radiating artificial radionuclides

no more than n·1 - n·10

Limit of the main permissible error of measurements, %

no more than ±30

Range of measurement of activities, Bq

- 137Cs

5 - 20·103

- 90Sr+90Y

1 - 2·103

- 226Ra

12 - 20·103

- 232Th

8 - 8·103

- 40K

50 - 20·103

- other gamma-radiating artificial radionuclides

n·1 - n·106

Power, from the AC mains with frequency 50 Hz, V

220

Power consumption, VA, no more than

200

Range of operating temperatures, oC

10 - 35

Weight, kg

- of detection units BDEG and BDEB

3,4

- of analog-to-digital converter MD-198

0,4

- of BDEG-K unit with Dewar flask and liquid nitrogen

40

- of low-background chamber with passive shielding for BDEB unit

80

- of low-background chamber with passive shielding for BDEG unit

200

- of low-background chamber with passive shielding for BDEG-K unit

200

- of spectrometer-radiometer MKGB-01 in assembly
(with units BDEG-63/80/150)

300

Overall dimensions of shielding units, mm

- BDEG

450х600

- BDEG-K with Dewar flask

480х800

- BDEB

300х400

Capacity of container for measuring materials, l

- Marinelly container

1

- Cylindrical container ("mayonez")

0,25

- Cuvette

0,038

MKGB-01

Purchase equipment

Изображение 1. Spectrometers and radiometers of radiation : Gamma and beta spectrometer - radiometer MKGB-01 RADEK Изображение 2. Spectrometers and radiometers of radiation : Gamma and beta spectrometer - radiometer MKGB-01 RADEK Изображение 3. Spectrometers and radiometers of radiation : Gamma and beta spectrometer - radiometer MKGB-01 RADEK Изображение 4. Spectrometers and radiometers of radiation : Gamma and beta spectrometer - radiometer MKGB-01 RADEK Изображение 5. Spectrometers and radiometers of radiation : Gamma and beta spectrometer - radiometer MKGB-01 RADEK